7.8 Replacing the press plunger

In order to facilitate replacing the press plunger, the following
procedure is recommended:

1. Remove the screw (20) and press drive cover (56) while the

furnace head is closed.

2. Loosen the terminal screw from the press plunger (65) by about

half a rotation.

3. Open the furnace head by means of the respective key (90). When

the furnace head is wide open, switch off the furnace,
disconnect the power plug and let the furnace cool to room
temperature.

4. Pull the press plunger (59) with slightly rotating movements from

the split taper socket (64) with one hand and pull down the press
plunger in the firing chamber with the other hand.

Contraindication:

Do not touch the thermocouple when replacing the press
plunger.

5. Push the white press plunger

(59) with the taper ahead into
the guide bush. Push the press
plunger with slightly rotating
movements into the split taper
socket (64) and fasten the screw
(65).

Contraindication:
Never reach into the press drive during

operation. There is a risk of crushing.

Mount the press drive cover (56) and fasten with screw.

6. Connect the power plug and switch on the furnace with the 0/I

switch.

20

56

65

64

59

7.6 Press power calibration