Product specifications, 1 delivery form, 2 technical data – Ivoclar Vivadent EP5000-G2 User Manual

Page 37: 3 acceptable operating conditions, 4 acceptable transportation and storage conditions

Advertising
background image

37

9. Product Specifications

9.1 Delivery form

– Programat EP 5000/G2
– Cooling tray “L”
– Programat Firing tray Kit
– Firing plate
– Automatic Temperatur Checking Set 2 ATK 2 (Test Set)
– USB download cable
– Programat USB stick
– PrograBase 2 Software
– Service passport equipment
– Operating Instructions
– Cooling grid

9.1.2 Recommended accessories
– Programat Accessories Set
– Automatic Temperatur Checking Set 2 ATK 2
– Vacuum Pump VP4

9.2 Technical Data

Power supply

110–120 V / 50–60Hz

200–240 V / 50–60Hz

Overvoltage category II
Contamination level 2

Tolerated voltage fluctuations

+/- 10%

Max. power consumption

12 A at 110–120 V

8.5 A at 200–240 V

Acceptable data for vacuum pump of other manufacturers
Max. output:

250 W / max. leakage current 0.75 mA

Final vacuum:

< 50 mbar

Use only tested pumps

Electrical fuses:

110–120 V:

250 V / T 15 A (heating circuit)

250 V / T 5 A (vacuum pump)


200–240 V:

250 V / T 8 A (heating circuit)

250 V / T 3.15 A (vacuum pump)

Dimensions of electrical fuses:

110–120 V:

Diameter 6.3 x 32 mm

200–240 V:

Diameter 5 x 20 mm

Dimensions of the closed furnace:

Depth: 470 mm / width: 305 mm / 400 mm (with cooling tray)

Height: 565 mm

Usable size of the firing chamber:

Diameter 80 mm

Height 48 mm

Max. firing temperature :

1200 °C

Weight:

Furnace base: 13.5 kg

Furnace head: 7.0 kg

Safety information
The EP 5000/G2 complies with the following guidelines:
– EN61010-1:2001 2

nd

Ed.

– IEC61010-1:2001 2

nd

Ed.

– UL/CSA61010-1:2004 2

nd

Ed.

– EN61010-2-010:2003 2

nd

Ed.

– IEC61010-2-010:2003 2

nd

Ed.

– CSA61010-2-010:2004 2

nd

Ed

Radio protection / electromagnetic compatibility

EMC tested

9.3 Acceptable operating conditions

Acceptable ambient temperature range:
+5 °C to +40 °C (+41 °F to +104 °F)

Acceptable humidity range:
80 % maximum relative humidity for temperatures up to 31 °C
(87.8 ° F) gradually decreasing to 50 % relative humidity at 40 °C
(104 °F); condensation excluded.

Acceptable ambient pressure:
The furnace is tested for use at altitudes of up to 2000 m (6562 ft.)
above sea level.

9.4 Acceptable transportation and storage conditions

Acceptable temperature range

-20 to +65 °C (-4 °F to +149 °F)

Acceptable humidity range

Max. 80 % relative humidity

Acceptable ambient pressure

500 mbar to 1060 mbar

Use only original packaging of the Programat EP 5000/G2 together