Ivoclar Vivadent EP5000-G2 User Manual

Page 35

Advertising
background image

35

Please contact the Ivoclar Vivadent After Sales Service, if one of the
following error messages is being displayed:

25, 29
43, 44, 45, 46, 47, 48
54, 56
103, 107
143, 144, 145, 146, 147, 148
700, 701, 703, 704, 705, 706, 707
1010, 1011, 1012, 1013, 1014, 1015, 1016, 1017, 1018, 1019

1024, 1025, 1026, 1028
1143, 1144, 1145, 1146, 1147, 1148
1202, 1203, 1204, 1205, 1206, 1207
1300, 1301
1400, 1401, 1402
1500, 1530

*

Furnace head opens when this error occurs.

** A program in progress is stopped.
*** The error cannot be acknowledged; the programs cannot be started.

Error /

Hint

Nr.

Weiterar-

beit nach

Fehler

möglich

Fehler

Text Fehlermeldung

520

Error muffle crack

CDS Crack Detection System has been activated. The program has been aborted and the press

plunger has been moved backwards. CDS could probably save your restorations from muffle cracks.

Please check your restorations before you continue your working progress.

521

Error muffle crack

CDS Crack Detection System has been activated. The program has been aborted and the press

plunger has been moved backwards. CDS could probably save your restorations from muffle cracks.

Please check your restorations before you continue your working progress.

522

Error muffle crack

CDS Crack Detection System has been activated. The program has been aborted and the press

plunger has been moved backwards. CDS could probably save your restorations from muffle cracks.

Please check your restorations before you continue your working progress.

526

T is > B + 200°C at the start of

a IPF press progr.

Firing chamber is too hot for the start of a IPF press program.

527

IPF press program parameters

not plausible

The set IPF press program parameters are not plausible.

528

T is < 350°C at the start of a IPF

press program

The furnace is too deeply cooled down for the start of a IPF press program.

529

IPF press program Error

temperature

The temperature during the IPF press program does not comply with the specifications. The

program may be continued nonetheless.

530

Error during logging of press

program data

An error has occurred during logging of press program data. The storage medium might be full.

700

Supply voltage outside the

acceptable range

The supply voltage is outside the acceptable range. Check the supply voltage.

701

***

no

Start-up aborted due to an error The self-test of the furnace was interrupted by an error. It is not possible to work with the furnace!

Switch the furnace off and on again, once the error has been rectified.

702

Brief power failure during of a

program in progress

A program in progress was interrupted by a brief power failure. The program is continued!

703

Power failure during a program

from the memory stick in pro-

gress memory stick no longer

present

A program in progress (started from the USB memory stick) was interrupted by a power failure.

The program could not be continued, since the USB memory stick is no longer present!

704

Power failure during an over-

night program in progress

An overnight program in process was interrupted by a power failure. The program is continued!

707

Incorrect supply voltage

The furnace is operated with the incorrect supply voltage. Make sure that the furnace is operated

with the supply voltage indicated on the rating plate.

800

Final vacuum value not reached

The required final vacuum value cannot be reached. Check the vacuum pump.

801

Vacuum drop

An unacceptable vacuum drop has occurred.

802

The vacuum does not increase

(self-test)

No vacuum increase could be measured. Check the following points: Is the firing chamber tight

(no contamination on the sealing surfaces)? Is the vacuum hose connected? Is the vacuum pump

connected? Is the fuse F1 o.k.?

1302

**

ATK2 calibration:

Pre-heating to 962 °C

Error during calibration. Sample may not be correctly inserted. Try again with a new sample and

make sure the sample makes ample contact.

1310

Calibration reminder

Some time has passed since the last calibration procedure. Calibrate the furance soon.

1311

Calibration reminder – press

cycles

Some press cycles has been done since the last calibration procedure. Calibrate the furance soon.

1312

Dehumidification reminder

Some time has passed since the last dehumidification. Conduct a dehumidification in the near

future.

1510

Temperature > VT at the start

of a firing program

The temperature in the firing chamber is higher than the pre-drying temperature. Press START to

continue the program despite the error message.

1520

Software update: Invalid file

The file selected for the software update is invalid. The update cannot be conducted.

1521

Software update: Invalid version

The software to be installed is older or the same as the one installed on the furnace. The update

cannot be conducted

1522

Software update: Error during

update

An error has occurred during the software update. Do NOT switch off the furnace and try again. If

the error reoccurs, try conducting the update via the USB interface.

1550

Change operation mode

The operation mode has been changed! Make sure that the furnace runs at the new stand-by

temperature long enough before you start a program.

1700

Service note

Two years have passed or the heating muffle has been in use for more than 1200 firing hours since

the last inspection of the furnace. Ivoclar Vivadent therefore recommends that your furnace should

be inspected. Please refer to the Equipment Service Passport or the Operating Instructions for

further information. The interval until the next appearance of the note can be set in the Settings.