Calibration of a single reference point, 5 oper ating instr u ctions – HEIDENHAIN ND 2100G User Manual
Page 44

44
1 Operation
1
.5 Oper
ating instr
u
ctions
Calibration of a single reference point
A single measurement reference point can be calibrated for encoders
and fully-calibrated transducers.
To calibrate a single reference point
U
Press the MASTER soft key
U
Use the arrow cursor keys to select the desired channel
U
Position the probe against the reference surface
U
Use the numeric keypad to enter the reference value (zero or the
desired offset) into the Master Value data field
U
Press the ENTER key
U
Press the YES soft key to calibrate the measurement reference
A green dot will appear next to the actual value indicating that the
reference point is calibrated.
Channel 1 selected. Reference value of 0.
Press the YES soft key.
Channel 1 calibrated.
The Master type should be set to Mean in the Master
setup screen to perform a calibration of a single encoder
point. Refer to "Master setup screen" on page 111.