Emerson Process Management ControlWave XFC User Manual

Page 128

Advertising
background image

F-26 / Appendix F

CI-ControlWave XFC

Low Flow Cut Off - When the differential pressure drops below the low flow cut off value,
the flow rate will be set to zero. Default units are “inches of water.”

Orifice Diameter - Orifice diameter is entered here. Default units are “inches.”

Pipe Diameter - Diameter of the pipe is entered here. Default units are “inches.”
Orifice Constant - K (AGA3 1985) - Orifice constant is entered here.

Adjust Press. - Users enter Average Barometric Pressure here.

Diff. Pressure - Actual value of Differential Pressure (Inches H

2

O) are displayed here.


Static Pressure - Actual value of Static Pressure (psig) are displayed here.

Temperature - Actual value of Temperature (Deg. F) are displayed here.

Specific Gravity - Specific Gravity of the gas being measured is displayed here.

Figure F-18A - (1985 AGA3) Orifice Flow Equation Setup Web Page


FPV
- Supercompressibility Factor (FPV) is displayed here.

Base Temperature - Required or Contract Base Temperature is entered here (Deg. F).

Base Pressure - Required or Contract Base Pressure is entered here (psig).

The following outputs from the AGA3 calculation are displayed:

MSCF/H - Flow rate in thousands of standard cubic feet per hour

Low Flow Cut Off - Cutoff (if the DP drops below the low flow cut off value) or OK

C Prime - Orifice Flow Constant

Fb - Basic Orifice Factor

Advertising