Emerson Network Router User Manual

Page 22

Advertising
background image

16

ProLink

®

II Software for Micro Motion

®

Transmitters

Installation and Setup

Figure 2-8

Modbus/RS-485 temporary connection to the service port

Figure 2-9

HART/Bell 202 temporary connection to HART clips

Attach serial port or USB adapter
if necessary (not shown)

Service port terminals

Transmitter

RS-232 <-> 2-wire RS-485
Interface Converter

Notes:
• For Model 1700/2700, Model 2400S, and LF-Series field-mount transmitters, the service port is available at any

time.

• For Model 1500/2500, Series 3000, and LF-Series DIN rail transmitters, the service port is available during a

10-second interval after power-up. If a service port connection is made during this time, the ports will remain in
service port mode.

• Ensure that no other Modbus master devices are currently active on the network.
• All service ports are accessed using the default address of 111. If you are connecting over a multidrop network

with multiple service ports, it is not possible to specify which device to connect to.

VIA
TOR

HART clips

Bell 202 signal

converter

Notes:
• If necessary, add resistance across the HART clips to bring the total resistance within 250–600

Ω.