2 acknowledging alarms, 10 managing the prolink ii license, Ense (see section 3.10) – Emerson Network Router User Manual

Page 40

Advertising
background image

34

ProLink

®

II Software for Micro Motion

®

Transmitters

Using ProLink II Software

Figure 3-7

Alarm Log window

3.9.2

Acknowledging alarms

To acknowledge an alarm using ProLink II, you must use the

Alarm Log

window.

To acknowledge an alarm:

1. Choose

ProLink > Alarm Log

.

2. Select

Ack

for each alarm you want to acknowledge.

-

If the alarm is no longer active (green light), the alarm will be cleared from the list, and the

Alarm Log

window will be updated as soon as new data is received from the transmitter.

-

If the alarm is still active (red light), the red light will remain, but the name of the alarm
will be grayed out. The alarm will be cleared from the list as soon as it is no longer active.

3.10

Managing the ProLink II license

Both the ProLink II temporary license and permanent license are keyed to a specific disk and a
specific folder or directory on a specific PC. If you want to move the ProLink II installation, you must
use the appropriate license transfer function:

To Directory

is used to transfer ProLink II to another location on the same PC.

Out of Computer

and

Into Computer

are used to transfer ProLink II to another PC.