Emerson Network Router User Manual

Page 25

Advertising
background image

Installation and Use Manual

19

Installation and Setup

Pr
oL
in

k II

Setup

T

ransmitter Star

tup

Usi

ng
Pr
oL
in

k

II

Be
fo

re
Y

o

u

Begin

If you use any other connection type, you must configure ProLink II connection parameters to
match the transmitter’s configuration. If you do not know the transmitter’s configuration, you
can use a Communicator or the transmitter’s display to view or change its configuration.

-

For all transmitters, if you are using HART protocol, you can specify the transmitter’s
HART tag (software tag) instead of the HART address, if a HART tag has been configured
in the transmitter.

-

For all Series 1000/2000/3000 transmitters, see Table 2-4 for default values for each
connection type.

-

For LF-Series transmitters, see Table 2-5 for default values for each connection type.

-

For IFT97xx transmitters, communication parameters are not configurable. Settings are
listed in Table 2-6. Configure ProLink II connection parameters to match these settings.

-

For RFT97xx transmitters, communication is configured using switches and jumpers on
the transmitter. Check your transmitter and refer to the transmitter manual to determine
your transmitter’s configuration, then configure ProLink II connection parameters to
match these settings. Factory default settings for these transmitters are listed in Table 2-6.

Table 2-2

MVD Direct Connect auto-detection limits

Parameter

Option

Protocol

Modbus RTU (8-bit)
Modbus ASCII (7-bit)

Baud rate

Standard rates between 1200 and 38,400

Parity

Even, odd, none

Stop bits

1, 2

Table 2-3

Model 2400S service port auto-detection limits

Parameter

Option

Protocol

Modbus RTU (8-bit)
Modbus ASCII (7-bit)

Address

Responds to both:

• Service port address (111)
• Configured Modbus address (default=1)

Baud rate

Standard rates from 1200 to 38,400

Stop bits

1, 2

Parity

Even, odd, none