42

ProLink

®

II Software for Micro Motion

®

Transmitters

Initial Transmitter Startup Procedures

Figure 4-4

Zeroing

4. Change the

Zero Time

if necessary. You will need to click

Apply

to send the new value to the

transmitter before performing the actual zeroing procedure.

5. If the Convergence Limit parameter is displayed, type a new convergence limit in the

Converg. Limit

box or accept the default value.

6. Click

Perform Auto Zero

. The meter will begin zeroing. The

Calibration in Progress

status

light will turn red.

7. Wait until the zero time has expired. At the end of this time:

If the

Calibration Failure

status light turns red, the zero procedure failed.

-

See your transmitter manual for troubleshooting procedures.

-

If desired, use the buttons in the dialog box to restore the previous zero value or the
zero value established during factory calibration.

Note: These two functions are not available on all transmitters.

If the

Calibration in Progress

status light returns to green and the

Calibration Failure

status light does not turn red, the zero procedure succeeded.

8. Click

Close

.