Initial Transmitter Startup Procedures
Figure 4-4
Zeroing
4. Change the
Zero Time
if necessary. You will need to click
Apply
to send the new value to the
transmitter before performing the actual zeroing procedure.
5. If the Convergence Limit parameter is displayed, type a new convergence limit in the
Converg. Limit
box or accept the default value.
6. Click
Perform Auto Zero
. The meter will begin zeroing. The
Calibration in Progress
status
light will turn red.
7. Wait until the zero time has expired. At the end of this time:
•
If the
Calibration Failure
status light turns red, the zero procedure failed.
-
See your transmitter manual for troubleshooting procedures.
-
If desired, use the buttons in the dialog box to restore the previous zero value or the
zero value established during factory calibration.
Note: These two functions are not available on all transmitters.
•
If the
Calibration in Progress
status light returns to green and the
Calibration Failure
status light does not turn red, the zero procedure succeeded.
8. Click
Close
.