Figure 5-2
Sample calibration tags – T-Series sensor
Figure 5-3
Sample calibration tags – All sensors except T-Series
Density calibration factors
If your sensor tag does not show a D1 or D2 value:
•
For D1, enter the Dens A or D1 value from the calibration certificate. This value is the
line-condition density of the low-density calibration fluid. Micro Motion uses air.
•
For D2, enter the Dens B or D2 value from the calibration certificate. This value is the
line-condition density of the high-density calibration fluid. Micro Motion uses water.
If your sensor tag does not show a K1 or K2 value:
•
For K1, enter the first 5 digits of the density calibration factor. In the sample tag in Figure 5-3,
this value is shown as
12500
.
•
For K2, enter the second 5 digits of the density calibration factor. In the sample tag in
Figure 5-3, this value is shown as
14286
.
If your sensor does not show an FD value, contact Micro Motion customer service.
If your sensor tag does not show a DT or TC value, enter the last 3 digits of the density calibration
factor. In the sample tag in Figure 5-3, this value is shown as
4.44
.
Newer tag
Older tag
Newer tag
Older tag
19.0005.13
19.0005.13
0.0010
0.9980
12502.000
14282.000
4.44000
310
12500142864.44
12500142864.44