Res 5-2 and 5-3, Density calibration factors – Emerson Network Router User Manual

Page 54

Advertising
background image

48

ProLink

®

II Software for Micro Motion

®

Transmitters

Transmitter Configuration, Characterization, and Calibration

Figure 5-2

Sample calibration tags – T-Series sensor

Figure 5-3

Sample calibration tags – All sensors except T-Series

Density calibration factors

If your sensor tag does not show a D1 or D2 value:

For D1, enter the Dens A or D1 value from the calibration certificate. This value is the
line-condition density of the low-density calibration fluid. Micro Motion uses air.

For D2, enter the Dens B or D2 value from the calibration certificate. This value is the
line-condition density of the high-density calibration fluid. Micro Motion uses water.

If your sensor tag does not show a K1 or K2 value:

For K1, enter the first 5 digits of the density calibration factor. In the sample tag in Figure 5-3,
this value is shown as

12500

.

For K2, enter the second 5 digits of the density calibration factor. In the sample tag in
Figure 5-3, this value is shown as

14286

.

If your sensor does not show an FD value, contact Micro Motion customer service.

If your sensor tag does not show a DT or TC value, enter the last 3 digits of the density calibration
factor. In the sample tag in Figure 5-3, this value is shown as

4.44

.

Newer tag

Older tag

Newer tag

Older tag

19.0005.13

19.0005.13

0.0010
0.9980

12502.000
14282.000

4.44000

310

12500142864.44

12500142864.44