B.3.1 flow source, B.3.2 control options, Section b.3.1) and click – Emerson Network Router User Manual
Page 82: Ee section b.3.2)

76
ProLink
®
II Software for Micro Motion
®
Transmitters
Configuring the Discrete Batch Application
B.3.1
Flow source
The flow source specifies the flow variable that will be used for batch measurement. Select one of the
flow sources defined in Table B-1.
B.3.2
Control options
The batch control options are used to manage the batch process. Control options are listed and defined
in Table B-2.
Note: Control options apply to all batch presets.
Table B-1
Flow sources
Flow source
Description
None
• Batch controller is disabled.
• START button will not appear on display.
Frequency input
• Frequency input from a Micro Motion IFT9701 or RFT9739 transmitter
• Frequency input from a pulse output device
Mass flow rate
Mass flow rate from Series 3000 transmitter
Volume flow rate
Volume flow rate from Series 3000 transmitter
CM std vol flow rate
• Standard volume flow rate at reference temperature
• Standard volume flow is available only if the concentration measurement
application software is installed and configured to indicate standard volume
flow.
CM mass flow rate
• Net mass flow rate
• Net mass flow is available only if the concentration measurement application
software is installed and configured to indicate net mass flow.
CM vol flow rate
• Net volume flow rate at reference temperature
• Net volume flow is available only if the concentration measurement application
software is installed and configured to indicate net volume flow.
API temperature-corrected
volume flow
• Volume flow adjusted by the calculated volume correction factor
• Available only if the petroleum measurement application is installed and
enabled.
Table B-2
Control options
Setting
Default
Description
Enable
batch
Yes
• Select Yes to enable the discrete batch application.
• Select No to disable the discrete batch application. The operation mode will
default to the process monitor.
Count up
Yes
• If set to Yes, the total displayed on screen increases from zero to the target
value.
• If set to No, the total displayed on screen decreases from the target value to
zero
• The setting of the Count up control option affects only the quantity displayed
on screen. It does not affect configuration of presets.
• If the custody transfer application is installed, Count up is set to Yes and
cannot be changed.
Enable AOC
Yes
• Select Yes to enable Automatic Overshoot Compensation (AOC).
• When batch AOC is enabled and batch AOC calibration has been performed,
the batch controller compensates for the time required to close the valve.
• If Enable AOC is set to Yes, batch AOC calibration is required to provide data
for the compensation process. To perform batch AOC calibration, see
Section B.5.