Ac/dc voltage measurement – Amprobe Telaris-ISO-PRO Insulation-Tester User Manual

Page 21

Advertising
background image

AC/DC Voltage Measurement

Connected loads generating EMV disturbance or harmonic waves may impair the display
quality.

1)

Select the ”UAC/DC" measurement range using the measurement function selection
dial (19).

2)

Connect the red test lead to the "UAC/DC" socket and the black test lead to the
"COM" socket.

3)

Connect the test leads to the UUT.

4)

Select between AC or DC voltage pressing the ”AC/DC" key (17).

• When switching on or changing to the "UAC/DC" measurement function, the default

setting is always "AC".

5)

Read the measurement result on the display screen.

Overvoltage Protection for Voltage Measurement

If a external voltage of maximum 600 V AC/DC is connected by mistake to the UAC/DC
measurement socket, ISO-PRO will not be damaged. The instrument displays ">600V”. After
removal of the overvoltage wait for approx. 20 s to obtain correct measurement values.

• If an AC voltage >25 V is measured within the DC voltage range, the instrument displays

"Uext”.

• This is also valid for the AC range if measuring a DC voltage > 25 V.

19

PAEB30872322:Layout 1 02.10.2008 15:56 Uhr Seite 19