Installation, Operation, and Maintenance Manual # 509252 Rev R

105

Air Pressure Verification

To test air pressure inside the irradiator, remove one cover plate screw and, using a
magnehelic, place the end of the tube over the screw hole (see

Figure 43

). For the

I310M/I311M and I310M-D/I311M-D, the pressure should be 447-596 Pa (1.8-2.4”
H

2

O) with respect to outside (ambient) air.

For the I310MB irradiator (modular blower), pressure should be 447 - 822 Pa (1.8 to
3.3” H

2

O).

Figure 43. Air Pressure Test Points

Air Pressure Switch Replacement

1. Remove the irradiator cover.

HV Connector (J11)

Cover Plate

Cover
Screws