Installation, Operation, and Maintenance Manual # 509252 Rev R

113

Test Point Voltages.

Control Card Removal/Replacement

This unit must be returned to Fusion UV Systems for repair or direct replacement. The
control card cannot be repaired in the field.

1. Remove the six screws from the power supply cover.

2. Remove the cover.

WARNING:

Before taking magnetron current

readings, turn the unit off. Check resistance

between “MAG CURRENT” test points.

Reading should be approximately 40K Ohms.

If this value is not correct, do not attempt

current reading. Damage could result to the

meter or personnel injury could result. Contact

Fusion UV Systems Customer Service if 40K

Ohms reading is not obtained.

After checking resistance, be sure to switch

the meter to correct setting for DC voltage

reading.

Table 21. Operating Voltages

Test Points

Status

Standby

Lamp On

Photoresistor Test (TP101 and TP102)

12 VDC

0-6 VDC

Magnetron Current Test MAG A - COM
(TP105 and TP106)

0

0.650-0.750 VDC

WARNING:

Disconnect all power sources

before servicing high voltage components.