Test point voltages, Table 19. operating voltages, Control card removal/replacement – Fusion F300S User Manual

Page 107

Advertising
background image

Installation, Operation, and Maintenance Manual # 509252 Rev R

107

Test Point Voltages

Control Card Removal/Replacement

This unit must be disconnected and returned to Fusion UV Systems for repair or direct
replacement. The control card cannot be repaired in the field.

1. Remove the six screws from the power supply cover. Remove the cover.

2. Remove all connectors. Remove three screws at the top of the card. Slide the

card approximately an inch away from the front panel and raise it up out of its
bottom track.

3. To install a replacement card, reverse this procedure. Note the location of

connectors in

Figure 44 on page 109

and reinstall as shown.

WARNING:

Before taking magnetron current

readings, turn the unit off. Check resistance

between “MAG CURRENT” test points.

Reading should be approximately 40K Ohms.

If this value is not correct, do not attempt

current reading. Damage could result to the

meter or personnel injury could result. Contact

Fusion UV Systems Customer Service if 40K

Ohms reading is not obtained.

After checking resistance, be sure to switch

the meter to correct setting for DC voltage

reading.

Table 19. Operating Voltages

Test Points

Application

Status

Standby

Lamp On

Photoresistor Test (TP101 and TP102)

P300M/ P300MQ

12 VDC

0-6 VDC

Magnetron Current Test MAG A - COM
(TP105 and TP106)

P300M/P300MQ

0

0.650-0.750 VDC