Installation, Operation, and Maintenance Manual # 509252 Rev R

71

Switch from
OFF to
STANDBY

BLWR
PRESSURE
indicator lights

Single-phase
voltage missing
phase

See

“Basic Checks” on page 65

.

Air pressure

Measure air pressure inside irradiator
following the procedure described in

Chapter 6, “Maintenance”

. IF PRESSURE

IS LOW:

Dirty air filters

Replace dirty air filter on remote or
modular blower.

Pressure switch
faulty

Replace the pressure switch.

Air handling design

Check for adequate air handling
requirements.

STANDBY
switch does
not light

HV cable
disconnected

Connect HV cable.

No fault
indicators light.

Faulty HV cable

Inspect cable, check for continuity.

Control card

Replace control card.

LAMP OUT
and/or PWR
IMBAL
indicators light
immediately

Card connectors not
secure

Check control card. All connectors must be
properly secured.

Control card

Replace control card.

RF INTLK
indicator lights
or flashes
alternately with
the lamp
control
switches

No RF detector
connected

Connect RF detector. A system must have
at least one RF detector connected.

Programming error:
1. All units have
S1-5 open.
2. Unit(s) with S1-5
open and no jumper
between J105A-3
and J105A-4.

1. Close S1-5 on unit with RF detector
connected.

2. Install jumper.

RF cable shorted or
open

Check cable for continuity.

RF detector faulty

Replace RF detector.

Control card

Replace control card.

Table 18. Troubleshooting/Fault Diagnosis (continued)

Operation

Signal/

Other

Symptoms

Possible Cause

Corrective Action