F300S/F300SQ Ultraviolet Lamp System

Switch from
STANDBY to
ON
(continued)

PWR IMBAL
indicator lights

Power Supply

Usually, a diode or capacitor failure will
reduce magnetron current by one-half.
Test the diodes and capacitors according
to

Chapter 6, “Maintenance”

. The voltage

to the primary coils ONLY of the high
voltage transformers should also be
checked.

Filament
Transformer

Check the filament transformer and
connections.

Control Card

Replace the control card if no power
supply, magnetron, or other malfunctions
are found.

Failing (Moding)
Magnetron

Check magnetron current at front panel
test points. If it is not within specifications,
replace magnetron.

PWR ERROR
indicator lights

Filament
transformer

Check filament transformer and
connections.

Magnetron

Check for presence of magnetron current.
If it is zero, disconnect all power, and
replace magnetron. If magnetron current is
low or half, check diode block and HV
capacitors.

Control Card

Replace the control card.

RF INTLK
indicator lights

Lamp bulb faulty

Replace the lamp bulb.

RF Screen
damaged

Replace the RF screen.

RF - arcs forming

Inspect reflector for pitting; replace if
necessary.

RF Detector faulty

See

“RF Detector” on page 20

to test RF

detectors.

Position

Reposition the detector farther from the
screen and/or the light shield opening.

WARNING: The RF detector should at all times be closer to the screen or light shield
opening than personnel working nearby. It should be aimed so that it intercepts RF
radiation from the irradiators.

Table 18. Troubleshooting/Fault Diagnosis (continued)

Operation

Signal/

Other

Symptoms

Possible Cause

Corrective Action