F300S/F300SQ Ultraviolet Lamp System

Switch from
STANDBY to
ON
(continued)

LAMP OUT
indicator lights
after 8
seconds;
Lamp bulb
does not light
(continued)

Diodes, capacitors

If the readings are not different, a power
supply fault is indicated. Disconnect all
power and inspect for wiring damage. Test
diodes and capacitors according to the
procedures in

Chapter 6, “Maintenance”

.

Irradiator

Disconnect all power. Remove the
irradiator cover (see

Chapter 6,

“Maintenance”

). Inspect for obvious

damage to wiring. Inspect the reflector for
pitting near the large slots, which is
evidence of arcing.

Magnetron

If no power supply fault can be found,
disconnect all power. Replace magnetron
(see

Chapter 6, “Maintenance”

).

WARNING: If no magnetron test voltage is measured at the magnetron test points,
remove power from the unit immediately. Consult Fusion UV Systems Technical
Services.

Table 18. Troubleshooting/Fault Diagnosis (continued)

Operation

Signal/

Other

Symptoms

Possible Cause

Corrective Action