Installation, Operation, and Maintenance Manual # 509252 Rev R

59

NOTE: Times indicated are approximate and will vary. The exact interval is
determined by conditions within the system.

If additional RF detectors are connected to slaves, an RF fault in one of these
slaves will automatically be interpreted as a system fault shutting down the entire
system.

In a Master/Slave system, the Slave units must be switched to STANDBY or ON
before Master power supply is switched from OFF to STANDBY.

Remote Operation

The F300S/F300SQ can be operated remotely with a STANDBY/LAMP ON switch
connected to the Master. Refer to

Chapter 3, “Electrical System”

for instructions on

connecting remote switches.

The following procedure describes how to operate a lamp remotely:

1. Turn all remote switches OFF (open contacts).

Table 12. Routine Operation

Time

Power

Switch

Lamp

Control

Lamp Control

Flasher

Operation

0 sec

ON

Lamp Off

Red

Control card energized

1-2 sec

Standby

Yellow flash

Power supply FAN ON;
Modular Blower ON

10 sec

Steady yellow

Interlocks and fault detection
circuits function. Ready to
start.

0 sec

Lamp On
from
Standby

Green flash

Start HV transformers

2-8 sec

UV bulb ignites

8-10 sec

Steady green

UV bulb at full power

0 sec

From LAMP
OFF to
LAMP ON

Slow flash yellow

10 sec

Fast flash green

12-18 sec

UV bulb ignites

18-20 sec

Steady green

Full power

0 sec

From LAMP
ON to
STANDBY

Yellow flash

UV bulb extinguished

18 sec

Steady yellow

Ready to restart