Table 18. Troubleshooting/Fault Diagnosis

Operation

Signal/

Other

Symptoms

Possible Cause

Corrective Action

POWER switch
from OFF to ON

LAMP
OFF/RESET
button does
not light at
power up

Fuses F1 and F2

Check fuse and replace if blown.

Power supply wiring

If fuse immediately blows again, replace
fuses. Then: A. Disconnect P104 on back
of power supply, P10 from the irradiator,
and P20 at the RF detector. Press
STANDBY LAMP CONTROL button. If
fuse does not blow, proceed to (B). If the
fuse blows, the problem is in the internal
wiring or control card.

Control card

Replace control card.

Irradiator cable

B. Turn OFF. Reconnect P104. Press
STANDBY. If the fuse does not blow,
proceed to (C). If the fuse blows, the
problem is in the connectors or cable.

RF detector

C. Turn OFF. Reconnect P20 at the RF
detector. Press STANDBY. If fuse does not
blow, proceed to (D). If the fuse blows, the
problem is in the RF detectors or cable.

Filament
transformer

D. Turn OFF. Reconnect P10 to the
irradiator. Press STANDBY. If the fuse
blows, the problem is in the filament
transformer or connections.

Master turns
slaves ON but
not OFF

Programming error

Close S1-1 (YES) on master.

RF INTLK
indicator lights
on slave with
slave OFF,
master ON

Cable connection

Connect cable H760 to slave.

RF INTLK
indicator lights
on all units

Programming error

Close S1-5 (YES) on master.

System shuts
down; no fault
indicators light

Momentary loss of
AC

Turn off all units and then turn each unit
on. A momentary loss of power to one or
more units may be interpreted as a system
fault.