9 calculate<chn>:convert:snumber? 96, 10 calculate<chn>:convert:snumber <serl> 96, 11 calculate<chn>:convert:test? <res>|<volt> 96 – Fluke THE HART 1529 User Manual

Page 102: Calculate<chn>:convert:snumber, Calculate<chn>:convert:snumber <serl, Calculate<chn>:convert:test? <res>|<volt

Advertising
background image

CALC:CONV:PRIN 2

“PRB”,CHAN2,”ser”,090105,”TYP”,RES,”RANGE”,0

If no parameters are specified, the ALL parameter is assumed and the parame-
ters for all channels are printed to the current port. Specifying the <chn> pa-
rameter prints only the parameters for the specified channel (1 to 4) to the
current port. Optionally, the parameters may be printed to a specific port by
specifying the <port> parameter (see Table 10 on page 125, Port Numbers)

8.4.4.9

CALCulate<chn>:CONVert:SNUMber?

Returns the probe serial number for the specified channel.

CALC2:CONV:SNUM?

A_336C

The CALCulate suffix, <chn>, specifies the channel (1 to 4) or memory chan-
nel (5 to 14). The returned serial number is in string format.

8.4.4.10

CALCulate<chn>:CONVert:SNUMber <serl>

Sets the probe serial number for the specified channel.

CALC2:CONV:SNUM A_336C

The CALCulate suffix, <chn>, specifies the channel (1 to 4) or memory chan-
nel (5 to 14). The <serl> parameter is in string format. The <serl> parameter
can consist of up to eight characters that include any letters, numeric digits, and
the underscore ‘_’.

This command may be password-protected. To access this command:

• If the SYST:PASS:CONV? command returns a 0 (probe password-protec-

tion is disabled), the password is not required.

• If the SYST:PASS:CONV? command returns a 1 (probe password-protec-

tion is enabled), the password must be successfully entered using the
SYST:PASS:CEN <pass> command (see Section 9.4.11, Password Com-
mands).

8.4.4.11

CALCulate<chn>:CONVert:TEST? <res>|<volt>

Tests the temperature conversion for the specified channel.

CALC2:CONV:TEST? 100.0145

0.0100

The CALCulate suffix, <chn>, specifies the channel (1 to 4) or memory chan-
nel (5 to 14). For PRT/Thermistor channels, the <res> parameter specifies a re-
sistance value to convert to temperature. The <res> parameter is always in
Ohms for PRTs and KOhms for Thermistors. For Thermocouple channels, the
<volt> parameter specifies a voltage value to convert to temperature. The

1529 Chub-E4 Thermometer Readout
User’s Guide

96