3 specifications and environmental conditions 9, 1 specifications 9, Specifications 9 – Fluke THE HART 1529 User Manual

Page 18: Specifications and environmental conditions, Specifications, 3specifications and environmental conditions, 1 specifications

Advertising
background image

3

Specifications and Environmental
Conditions

3.1

Specifications

PRT

Thermistor

Thermocouple

Inputs

2 channels PRT/thermistor and 2 channels thermocouple
or 4 channels PRT/thermistor
or 4 channels thermocouple
PRT/thermistor channels accept 2,3, or 4 wires
Thermocouple channels accept B, E, J, K, N, R, S, T, and Au-Pt thermocouple types

Temperature Range

1

–189 to 960°C

–50 to 150°C

–270 to 1800°C

Measurement Range

0 to 400

Ω

0 to 500k

Ω

–10 to 100 mV

Resistance / Voltage
Accuracy

2

0 to 20

Ω : ±0.0005Ω

20 to 400

Ω: ±25 ppm of reading

0 to 5k

Ω: ±0.0005kΩ

5 to 200k

Ω: ±100 ppm of

reading
200 to 500 K

Ω: ±300 ppm of

reading

–10 to 50 mV: ±0.005 mV
50 to 100 mV: ±100 ppm of
reading
(Internal RJC: ±0.25°C)

Characterizations

ITS-90,
PT-100 (ASTM E1137, DIN 43760,
IEC-751),
Callendar-Van Dusen

Steinhart-Hart, YSI-400

NIST Monograph 175, 3-point
deviation function applied to
NIST 175, 6th-order polynomial

Temperature Accuracy (meter
only)

1

±0.004°C at –100°C
±0.006°C at 0°C
±0.009°C at 100°C
±0.012°C at 200°C
±0.018°C at 400°C
±0.024°C at 600°C

±0.0025°C at 0°C
±0.0025°C at 25°C
±0.004°C at 50°C
±0.010°C at 75°C
±0.025°C at 100°C

Type Ext./Int.RJC
B

±0.6°C/±0.6°C

E

±0.07°C/±0.25°C

J

±0.1°C/±0.35°C

K

±0.15°C/±0.4°C

N

±0.15°C/±0.3°C

R

±0.4°C/±0.5°C

S

±0.5°C/±0.6°C

T

±0.1°C/±0.3°C

Temperature Resolution,
typical

3

0.001°

0.0001°

0.01 to 0.001°

Operating Range

1

16 to 30°C

Measurement Interval

0.1 second to 1 hour; inputs may be read sequentially or simultaneously at 1 second or greater interval

Excitation Current

1mA
(reversing)

2 and 10

μA, automatically

selected

N/A

Display

1.3” x 5" backlit LCD graphical display

Display Units

°C, °F, K,

Ω, KΩ, mV

Data Logging

Up to 8,000 time- and date-stamped measurements can be logged

Logging Intervals

0.1, 0.2, 0.5, 1, 2, 5, 10, 30, or 60 seconds; 2, 5, 10, 30, or 60 minutes

Averaging

Moving average of most recent 2 to 10 readings, user selectable

Statistical Data

Average, standard deviation, minimum, maximum, spread, N, difference (delta)

9

3 Specifications and Environmental Conditions

Specifications