channel, the probe serial number and the conversion type. The characterization
coefficients and parameters are displayed after the conversion type is selected.

The PROBE parameter is for selecting any one of the four input channels:
CHAN1, CHAN2, CHAN3 and CHAN4 or one of the ten probe memory chan-
nels: PMEM0, PMEM1, PMEM2, PMEM3, PMEM4, PMEM5, PMEM6,
PMEM7, PMEM8, and PMEM9. The probe memory channels are for storing
probe settings that can later be used to copy to an input channel.

The SERIAL parameter is the serial number of the probe being edited. The se-
rial number consists of a string of up to eight characters using any numeric dig-
its, letters, and underscore.

The TYPE parameter is the conversion type for the corresponding input chan-
nel. The list of possible conversion types for each probe type is given in Table
2. The conversion types available are dependent on the probe or channel type as
indicated in the table. The characterization coefficients and parameters are dis-
played after the conversion type is selected. These coefficients may change de-
pending on the selected conversion type. The conversions and related
coefficients and parameters are explained in the following sections.

The

LR buttons are used to select the parameter setting. The UD buttons

are used to scroll between and to change each digit of the parameter. Press the
Enter button to change and save the new setting. Press and hold the Exit button
to cancel to the main display or press the EXIT button to cancel and to move to

35

7 Menu Functions

Probe Menu

1

24.159C

3

24.142C

2

24.634C

4

24.015C

URANGE:

100

Ω

PROBE:

WIRES:

4

SERIAL:

145-560

RO:

100.0000

TYPE:

CVD

DAL:

0.003850000

CHAN-1

EXIT

ENTER / MENU

UNIT

CONTRAST

1529

CHUB E-4 THERMOMETER READOUT

Figure 14 Edit Probe