3 test probe 43, 4 print probe 43, Figure 16 test probe 43 – Fluke THE HART 1529 User Manual

Page 49: Print probe 43, Test probe 43, Test probe, Print probe, Figure 16, 159c, 142c

Advertising
background image

7.2.3

Test Probe

The TEST PROBE function allows you to test the probe characterization algo-
rithm and characterization coefficients for a specific probe. The probe and
probe type must be setup in the EDIT PROBE function prior to entering the
TEST PROBE function. When this function is selected, the bottom portion of
the display shows the probe channel or memory location, the probe serial num-
ber, and the conversion type.

Use the

LR buttons to select the channel or memory location. After the

channel is selected, press the Enter button to test the temperature calculation
for the probe. You can enter an input value (

Ω, KΩ, or millivolts) and the corre-

sponding output value (e.g. temperature) is shown. Press the Exit button to exit
to the menu.

7.2.4

Print Probe

The PRINT PROBE function allows you to print probe characteristics to the se-
rial port. Probe characteristics for a specific channel or memory location can be

43

7 Menu Functions

Probe Menu

1

24.159C

3

24.142C

2

24.634C

4

24.015C

PROBE:

CHAN1

SERIAL:

145-560

TYPE:

CVD

EXIT

ENTER / MENU

UNIT

CONTRAST

1529

CHUB E-4 THERMOMETER READOUT

Figure 16 Test Probe