3 channel mode 31, 4 moving average 31, Figure 10 channel mode 31 – Fluke THE HART 1529 User Manual

Page 37: Average 31, Channel mode 31, Filter 31, Moving average 31, Scan 31, Channel mode, Moving average

Advertising
background image

hold the Exit button to cancel to the main display or press the EXIT button to
cancel and to move to the next parameter.

7.1.3

Channel Mode

The CHANNEL MODE function sets the channel display mode. When this
function is selected, the bottom portion of the display shows the current scan
mode and allows the setting to be changed.

The following scan modes are available:

• SCAN - measures enabled channels scanning through and displaying the

channels one at a time, sequentially, in large digit format.

• SIMULT - measures and displays enabled channels simultaneously.

The

LR buttons are used to select a setting. Press the Enter button to save

the setting and exit. Press the Exit button to cancel and to exit to the menu.

If the measure period is 0.1 or 0.2 seconds, the CHANNEL MODE function is
unavailable. The channel mode is automatically set to scan mode. The follow-
ing message is displayed when the CHANNEL MODE function is selected.

NOT AVAILABLE AT THIS RATE.

Press ENTER to continue…

If the measure period is 0.5 seconds or greater, the channel mode operates
normally.

7.1.4

Moving Average

The MOVING AVERAGE function sets the input averaging. The moving aver-
age filter is useful for smoothing variations in the measurements and for im-

31

7 Menu Functions

Channel Menu

1

24.159C

3

24.142C

2

24.634C

4

24.015C

SC MODE:

SIMULT

EXIT

ENTER / MENU

UNIT

CONTRAST

1529

CHUB E-4 THERMOMETER READOUT

Figure 10 Channel Mode