The label suffix, <n>, specifies the number of the label (1 to 25) to retrieve.
The label name is returned in string format.

8.4.7.23

LOGging:LABel<n>:NAME <label>

Sets the name of the specified label.

LOG:LAB18:NAME BATH

The label suffix, <n>, specifies the number of the label (1 to 25) to set. Label
names can be up to 8 characters in length and can include any letters, numeric
digits, and the underscore (‘_’).

8.4.8

System Commands

The system commands are used for querying instrument identification parame-
ters, temperature units, and for setting operating parameters to the defined de-
fault values.

8.4.8.1

*IDN?

Returns the instrument identification string that indicates the manufacturer,
model number, serial number, and code firmware version.

*IDN?

HART,1529,A09001,1.11

8.4.8.2

*OPT?

Returns a list of input modules installed.

*OPT?

PRT,TC

Input module types are reported by position, top to bottom, of the two input
modules. The returned example is for a 1529 equipped with one 2-channel PRT
input module and one 2-channel Thermocouple input module.

8.4.8.3

*RST

Sets the instrument operating parameters to defined conditions.

*RST

The following parameters are set when this command is issued.

• Sets the scanning sequence timer to 1 second.
• Sets channel 1 ON and all others OFF.
• Sets the scan mode to SCAN.
• Sets the moving average to 1.
• Sets the printing to all ports OFF.