1 initiate 89, 2 initiate:continuous? 89, 3 sense:average:count? [min|max|def] 89 – Fluke THE HART 1529 User Manual

Page 95: 4 sense:average:count <num>|min|max|def 89, 5 sense<chn>:resistance:range? [min|max|def] 89, Initiate, Initiate:continuous, Sense:average:count? [min|max|def, Sense:average:count <num>|min|max|def, Sense<chn>:resistance:range? [min|max|def

Advertising
background image

8.4.2.1

INITiate

Provided for compatibility but has no effect.

INIT

8.4.2.2

INITiate:CONTinuous?

Provided for compatibility but has no effect.

INIT:CONT?

1

8.4.2.3

SENSe:AVERage:COUNt? [MIN|MAX|DEF]

Returns the moving average filter setting.

SENS:AVER:COUN?

4

Specifying the MIN parameter returns a value of 1, specifying the MAX pa-
rameter returns a value of 10, and specifying the DEF parameter returns a value
of 1. The *RST command sets the average count to the default (1).

8.4.2.4

SENSe:AVERage:COUNt <num>|MIN|MAX|DEF

Sets the moving average filter setting.

SENS:AVER:COUN MIN

The <num> parameter must be a number between 1 and 10. Specifying the
MIN, MAX, or DEF parameter sets the moving average filter setting to the
minimum, maximum, or default, respectively. The *RST command sets the
moving average to 1.

8.4.2.5

SENSe<chn>:RESistance:RANGe? [MIN|MAX|DEF]

Returns the resistance range setting.

SENS3:RES:RANG?

0

This command only applies to PRT/Thermistor channels. Executing this com-
mand on Thermocouple channels generates a -294, "Incompatible type", error
message.

The SENSE suffix, <chn>, specifies the channel (1 to 4), or memory channel (5
to 14). Specifying the MIN parameter returns a value of 0. Specifying the MAX
parameter returns a value of 1. Specifying the DEF parameter returns a value of
0. The *RST command sets the resistance range to the default (DEF).

89

8 Digital Communications Interface

Commands