The RANGE parameter is used to select the appropriate resistance range. For
25 ohm and 100 ohm PRTs select 100 ohms and for 1000 ohm PRTs select 10K
ohms.

The WIRES parameter sets the appropriate number of lead wires from the sen-
sor. This parameter can be set to 2-, 3-, or 4-wires. See Figure 5 on page 22 for
the wiring configuration.

The following table shows which parameter to set for each of the coefficients
that may appear on the certificate. Note: If the certificate has two sets of coeffi-
cients, one set for “zero-power” calibration and one set for 1 mA calibration,
use the coefficients for the 1 mA calibration.)

Matching Certificate Values to the 1529 ITS-90 Coefficients

1529 ITS-90 Coefficients

Certificate Value

A

a6, a7, a8, a9, a10, or a11

B

b6, b7, b8, or b9

C

c6, c7

D

d

A4

a4

B4

b4

The following example demonstrates how to set the ITS-90 parameters for the
case where a PRT was calibrated to ITS-90 and its calibration certificate states
values for coefficients Rtpw, a4, b4, a8, and b8. Set the 1529 parameters with
values from the certificate as follows.

Setting Coefficients Rtpw, a8, b8, a4, and b4

1529 ITS-90 Coefficients

Certificate Value

RTPW

Rtpw

A

a8

B

b8

C

0

D

0

A4

a4

B4

b4

7.2.1.2

ITS-SR5

The ITS-SR5 conversion is a special case of the ITS-90 conversion type. This
conversion is for the ITS-90 sub-range 5 only.

When the ITS-SR5 conversion type is selected, the RTPW, A5, and B5 coeffi-
cients are displayed on the right portion of the bottom of the display. The

UD

buttons are used to scroll through the coefficients. Use the Enter button to edit

37

7 Menu Functions

Probe Menu