3 view data 62, Figure 36 auto log view data 62, View data – Fluke THE HART 1529 User Manual

Page 68: Figure 36, Auto log view data, 159c, 142c, 634c, 015c

Advertising
background image

readings currently stored in the demand or auto log memory. The COUNT pa-
rameter shows the number of auto log readings remaining during the current
log session.

Press the Enter or Exit button to exit this display and to return to the AUTO
LOG menu.

Only those channels that are enabled during the log session are logged. Al-
though you may enable and disable channels during a log session we do not
recommend this practice. If channels are disabled during the log session, the
previously logged entries for those channels are not removed from the log.

From the main display, there are two ways to tell if data is being logged: (1) a
logging indicator in the lower left-hand corner (looks like a strip-chart) and (2)
by setting one of the display fields to the ALOG type (see Section 8.3.3, Edit
Fields)

7.4.2.3

View Data

The VIEW DATA function displays the label, date, reading number, channel,
measurement, units, and time of each auto log entry in sequential order. When
this function is selected, the bottom portion of the display shows the auto log
data.

Each time the view data function is accessed the current auto log memory is
displayed. The data is updated each time the function is accessed; it is not up-
dated continuously during a log session.

The entries are grouped by the label specified and displayed in the order that
they were logged. For each log session, an entry is written into the log, which
includes the label for all readings in that session and the date of the session.

1529 Chub-E4 Thermometer Readout
User’s Guide

62

1

24.159C

3

24.142C

2

24.634C

4

24.015C

DATA_01

10-12-00

0002:
0003:
0004:

12:23:24
12:23:24
12:23:24

24.147 C
29.637 C
23.972 C

1
2
3

EXIT

ENTER / MENU

UNIT

CONTRAST

1529

CHUB E-4 THERMOMETER READOUT

Figure 36 Auto Log View Data