2 start-up error messages 137, 2 downloading auto logged data 137, Calibration expired 137 – Fluke THE HART 1529 User Manual

Page 142: Memory 137, Start-up error messages, Downloading auto logged data, 2 downloading auto logged data, 2 start-up error messages

Advertising
background image

Message

Recommended Action

SYSTEMS: BATTERY NOT
INSTALLED

Cycle the power.
Check that the battery is properly connected.
Replace the battery.

CHANNELS: FAILURE

Cycle the power.

CALIBRATION: DUE FOR RECAL

Check that the instrument date is correct.
Recalibrate the instrument and set the recalibration date accordingly.

CALIBRATION: INVALID
PARAMETERS

Check the correctness of the calibration parameters for each channel.

GPIB: NOT INSTALLED

Normal message if the IEEE-488 option is not installed.

MEMORY DATA: QUESTIONABLE

Reset the memory by pressing EXIT when prompted.

MEMORY DATA: INVALID DATA

Reset the memory by pressing ENTER when prompted. (This action requires that important data
such as probe coefficients be restored.

BUTTONS: [button name]

Check the stated button to ensure that it is not stuck. Loosen as necessary.

11.1.2

Start-up Error Messages

The following messages may appear after the self-test. If the recommended ac-
tion fails to correct the error, contact an Authorized Service Center (see Section
1.3) for assistance.

Message

Recommended Action

CALIBRATION EXPIRED

Check that the instrument date is correct.
Recalibrate the instrument and set the recalibration date accordingly.

INVALID CALIBRATION PARAMETERS

Check the correctness of the calibration parameters for each channel.

MEMORY QUESTIONABLE

Reset the memory by pressing EXIT when prompted.

MEMORY DATA INVALID, MEMORY WILL BE
CLEARED

Reset the memory by pressing ENTER. (This action requires that important data
such as probe coefficients be restored.)

11.2

Downloading Auto Logged Data

The 1529 “Chub-E4” Thermometer Readout stores up to 8,160 autolog read-
ings or 100 demand log readings in its memory. These readings can be down-
loaded to a computer using the RS-232 serial port.

The following steps should be followed to download logged data using the se-
rial port:

1. Connect the 1529 to a COM port on the computer using the serial cable.

2. Run a terminal program such as Windows® Terminal or HyperTerminal®.

a) Set up the terminal program to use the correct COM port and baud rate. Con-
figure the communication settings to use 8 data bits, 1 stop bit, no parity and no
flow control.

137

11 Troubleshooting

Downloading Auto Logged Data

Advertising