Fluke TRACKERTM 90100 User Manual

Page 16

Advertising
background image

Model 90100
Operators Manual

2-4

WARNING

4. Power can now be applied to the system. Wait for the Test/Detector screen shown below before

pressing any keys.








Reverse the above procedure to disconnect the interface cable from the 35300A Detector.



Always turn the Model 35360A TRACKER Display's
power off prior to connecting or disconnecting the
interface cable. Failure to do so may result in an
electrical shock.

2.2.3 RS-232 Computer Interface (RJ-45 Style)

The RS-232 Computer Interface provides a means for computer controlled factory test and calibration. It
is also used in conjunction with the Customization program during the user customization process.

2.3 Positioning the Model 35300A Detector in the Beam

Proper and consistent positioning of the 35300A Detector is necessary for reliable measurement results.
The 35300A Detector has a printed overlay on both of its flat surfaces to indicate the location of the five
detectors inside the assembly. The overlays also include a black cross-hair to aid in centering the device
in the treatment beam.

The 35300A Detector should be placed so that the center detector will be in the center of the treatment
beam. For a gantry angle of 0 degrees, place the 35300A Detector on the treatment table with the axis
label face up, and the connector facing away from the gantry. The "TOP" chamber should be the one
nearest the gantry. This placement is shown in Figure 2-4.