Maintenance & Service

Troubleshooting

4

4-7

WARNING

Dose Rate Measurement Value Does Not Return to Zero in the Absence of a Radiation Beam

Figure 5-5

Dose Rate Measurement Zero Problem Flowchart

4.3 Line Fuse Replacement


Disconnect the 35360A TRACKER Display from
other equipment along with the AC line power and
turn off power before continuing with this section.

No

Yes

Switch 35360A TRACKER Display

to the dose measurement mode

and check for excessive leakage drift.

The problem appears to be caused by

one or more improper A-D Zero

setting. Use the TRACKER

Customization Software to readjust

the A-D Zero setting for each

drifting channel.

One or more dose rate measurement

values do not return to within one

count of zero in the absence of

radiation.

Dose measurement leakage drift

greater than 2 counts per minute?

The problem appears to be caused

by current leakage in one or more

measurement channels. Use the

Dose Leakage Troubleshooting chart,

Figure 5-4, to further diagnose the

problem.

When powering the 35360A TRACKER Display from

an AC line, always use a 3-wire grounding type line cord such

as the one supplied with the 90100 TRACKER System.

In addition, only connect the 35360A TRACKER Display to an

AC power outlet employing a third wire safety ground.

Always turn the 35360A TRACKER Display's power off

prior to connecting or disconnecting the interface cable.

Failure to do so may result in an electrical shock.

The Model 35360A TRACKER Display and 35300A

Detector contain no internal operator serviceable parts.

Disassembly of either unit could result in electrical shock.