Recommended build-up for photons is given in Table A-1; recommended build-up for electrons is given in
Table A-2. Table A-2 takes into account the 0.56 cm water-equivalent inherent in the 35300A Detector.
Cobalt-60 radiation does not require any added build-up.

Table A-1. Recommended Amount of Build-up to Use With the Model 35300A Detector to

Attain Photons d

max

for Linear Accelerators

Nominal

Accelerator Potential,

MeV

Number of build-up

plates

(37051C-1)

"Solid Water"

thickness (cm)

2 0 0

3 1 0.5

4 2 0.5

5 2 1.0

6 3 1.0

7 3 1.5

8 3 1.5

9 4 1.5

10 4 2.0

11 4 2.0

12 4 2.0

13 5 2.0

14 5 2.5

15 5 2.5

16 6 2.5

17 6 3.0

18 6 3.0

19 6 3.0

20 7 3.0

21 7 3.5

22 7 3.5

23 7 3.5

24 8 4.0

25 8 4.0