Figure 4-4. leakage flowchart – Fluke TRACKERTM 90100 User Manual

Page 42

Advertising
background image

Model 90100
Operators Manual


4-6

Dose Measurement Value Drifts Prior to (or following) the Exposure




















































Figure 4-4.

Leakage Flowchart

No

Yes

Yes

No

Yes

Yes

No

Yes

No

No

Yes

No

Clean 35360A TRACKER Display's

interface connector via procedure

given later in this section and

The drift problem has been

eliminated and the system is ready

for use.

Attempt to clean the 35300A Detectors's

interface connectors was unsuccessful.

35300A Detector will require

repair or replacement.

Excessive dose drift?

(35360A TRACKER Display,

35300A Detector and

The drift problem has been

eliminated and the system is ready

for use.

Clean 35300A Detector's interface

conncector via procedure given later in

this section and recheck dose drift.

Excessive dose drift?

Excessive dose drift?

(35360A TRACKER

Connect the 35300A Detector to the

interface cable and recheck complete

system leakage.

Attempt to clean interface cable

connectors was unsuccessful. Cable

will require repair or replacement.

Excessive dose drift?

(35360A TRACKER Display

Disconnect interface cable from

35360A TRACKER Display

One or more dose measurement values

exhibits a drift rate of greater than two

counts per minute in the absence of

radiation.

Clean interface cable connectors via

procudure given later in this section

and recheck dose drift.

Excessive dose drift?

(35360A TRACKER Display

Reconnect interface cable to the 35360A

TRACKER Display without the 35300A

Detector and recheck the dose drift.

Excessive dose drift?

(35360A TRACKER

Attempt to clean 35360A TRACKER

Display's interface connector was

unsuccessful. 35360A TRACKER

Display will require repair or replacement.

When powering the 35360A TRACKER Display from

an AC line, always use a 3-wire grounding type line cord such

as the one supplied with the 90100 TRACKER System.

In addition, only connect the 35360A TRACKER Display to an

AC power outlet employing a third wire safety ground.

Always turn the 35360A TRACKER Display's power off

prior to connecting or disconnecting the interface cable.

Failure to do so may result in an electrical shock.

The Model 35360A TRACKER Display and 35300A

Detector contain no internal operator serviceable parts.

Disassembly of either unit could result in electrical shock.

WARNING

and recheck dose drift.

Display only)

+ interface cable)

recheck dose drift.

Display only)

+ interface cable)

interface cable)

(35360A TRACKER Display,

35300A Detector, and

interface cable)