Before You Start- Safety Precautions

1

1-3

WARNING

CAUTION

1.3 Before You Start - Safety Precautions

This product is designed for quality assurance testing of therapeutic radiation treatment machines. It is
not for machine calibration nor is it intended as a substitute for a precision dosimeter system (Farmer-type
Probe and Electrometer with official calibration). Its use should be restricted to the monitoring of
constancy of output, flatness, and symmetry. It is the responsibility of the user to establish the correct
procedure of uses including the initial configuration of the device, (via the Customization program
included). Quality assurance testing should be done by qualified personnel under the supervision of a
trained medical physicist. The Model 90100 TRACKER System should be used only by people who are:

• aware of and understand the limitations of the device as they relate to the measurement of radiation

output;

• knowledgeable about safety procedures observed when working with radiation sources, such as

therapy radiation machines;

• aware of safety precautions required to avoid possible injury when using electrical/electronic

equipment.

Please take the time to read this manual carefully before operating this instrument.

1.4 Safety Terms

The following terms are used in this manual.


The

heading used in this manual explains dangers that might result in personal injury or

death. Always read the associated information very carefully before performing the indicated procedure.


The

heading used in this manual explains hazards that could damage the instrument.

Such damage may invalidate the warranty.

1.5 Manual Addenda

Any improvements or changes concerning the instrument or manual will be explained in an addendum
included with the manual. Be sure to note these changes and incorporate them into the manual.

1.6 Specifications

1.6.1 Model 35300A Detector

Source Modes and Energies

Photons: 300 keV to 25 MeV, with suitable build-up material

Electrons: 2 MeV to 25 MeV, with suitable build-up material