Model 90100
Operators Manual

2-12

NOTE





Figure 2-18. An Absolute Rate Measurement With Air Density Corrections and the Front Panel Gain

Factors Disabled


Prior to making a dose measurement, press the RESET button to zero all ion chamber readings.
Pressing this button has no affect when performing dose rate measurements.

2.4.8.2 Percent-of-Center Measurements

The Percent-of-Center Measurement screen displays the central ion chamber's reading in absolute
measurement units and the four peripheral ion chamber's readings as a percentage of the center ion
chamber's reading.

top

center

x 100%

left

center

x 100%

right

center

x 100%

bottom

center

x 100%

center units

Figure 2-19.

A Percentage Measurement Showing How the Ratios are Calculated







Figure 2-20. A Percentage Dose Measurement With Air Density Corrections and the Front Panel Gain

Factors Enabled

Figure 2-21.

A percentage rate measurement with both air density corrections and the front panel

gain factor disabled

Pressing the RESET button while performing a
dose measurement causes all percentages to
immediately change to "UNDER". This is normal
operation (see section 2.4.8.3 for more
information).