Appendix

Using Build-Up

A

A-3

Table A-2. Recommended Amount of Build-up to Use With the Model 35300A Detector to

Attain Electrons dmax for Linear Accelerators. 37051C-1 Plates are 0.44 cm Each

Nominal

Accelerator Potential,

MeV

Number of build-up

plates

(37051C-1)

"Solid Water"

thickness (cm)

2 1 0.5

3 1 0.5

4 1 0.5

5 2 1.0

6 2 1.0

7 3 1.5

8 3 1.5

9 4 2.0

10 5 2.0

11 5 2.5

12 6 3.0

13 6 3.0

14 7 3.5

15 7 3.5

16 8 4.0

17 8 4.0

18 9 4.5

19 10 4.5

20 10 5.0

21 11 5.5

22 11 5.5

23 12 6.0

24 12 6.0

25 13 6.0

The amount of underside build-up to be used depends on how closely you wish to approximate the back-
scattering of a water bath or other test set-up. Generally, several inches are used: about eight 37051C-1
plates (or about 4.0 cm of "Solid Water") regardless of the incident energy shown in the tables.