6 definitions – Fluke TRACKERTM 90100 User Manual

Page 51

Advertising
background image

Customization

Definitions

5

5-5

IMPORTANT

5.6 Definitions

This section describes all of the Items that may be customized with the Customization software. It is split
into two sections matching the two screens.

CUSTOMIZATION ITEMS

Power Down Interval

The power down interval is used to specify the number of minutes the 35360A TRACKER Display
remains powered during periods of inactivity. If the 35360A TRACKER Display senses no operator input
in the form of a key press, exposure, or RS-232 command for the duration of the power down interval, it
will turn itself off. This interval must be in the range of 5 to 255 minutes. An entry of zero will disable the
automatic power down feature.

Temperature Units

This Item provides a selection of temperature scales for use on the 35360A TRACKER Display's Air
Density screen. The choices are: Celsius, Fahrenheit, and Kelvin.

Pressure Units

This Item provides a selection of pressure scales for use on the 35360A TRACKER Display's Air Density
screen. The choices are: millimeters of mercury (Hg), inches of mercury, feet from sea level, meters from
sea level, millibars, and hectoPascals.

Optional Message

Text entered for this Item will appear on the bottom line of the 35360A TRACKER Display's Test screen.
The Item is limited to a maximum of 20 characters.

Ion Chamber Calibration Units (ICCU)

The Ion Chamber Calibration Units Item defines the units in which the ion chambers in the 35300A
Detector were calibrated. The default is roentgens per nanocoulomb.

This means that the Ion Chamber Calibration Factor (ICCF) given on the Display Calibration Items screen
MUST be in units of R/nC. Choices for units are roentgens per nanocoulomb, grays per nanocoulomb,
sieverts per nanocoulomb, and rads per nanocoulomb.

Display Units

The Display Units Item defines the units in which measurements will be displayed on the 35360A
TRACKER Display screen. The choices are roentgens, grays, sieverts, rads, and monitor units. The
default is rads.

Display Units Conversion Factor (DUCF)

The 35360A TRACKER Display uses the Display Units Conversion Factor (DUCF) Item to convert the
output of the ion chambers (in units defined by the Ion Chamber Calibration Units) to the units required for
display (defined by Display Units).



You must set the Display Units Conversion Factor
correctly and accurately to convert the ion chamber
output to displayed units. The default is 0.879
rads/R.