Maintenance & service troubleshooting, Figure 4-3. bias error flowchart – Fluke TRACKERTM 90100 User Manual

Page 41

Advertising
background image

Maintenance & Service

Troubleshooting

4

4-5

Bias Error (LoBias or HiBias) Indication at the Display

Figure 4-3.

Bias Error Flowchart

No

No

Yes

Yes

No

Yes

The 35300A Detector is faulty and

will require repair or replacement.

Interface cable is faulty and will require repair or

replacement.

The bias error has disappeared. Repeat the

troubleshooting procedure if the error indication

returns.

Bias error

annuciator displayed?

(35360A TRACKER Display,

35300A Detector, and

35360A TRACKER Display's electronic

bias circuit is faulty. 35360A TRACKER

Connect the 35300A Detector to the interface

cable and recheck VFD for bias error indication.

Disconnect interface cable from 35360A

TRACKER Display and recheck VFD

Bias error annuciator is displayed on

35360A TRACKER Display's VFD.

Bias error annuciator displayed?

(35360A TRACKER Display

Reconnect interface cable to the 35360A

TRACKER Display without the 35300A

Detector and recheck VFD for bias

Bias error annuciator displayed?

(35360A TRACKER

When powering the 35360A TRACKER Display from

an AC line, always use a 3-wire grounding type line cord such

as the one supplied with the 90100 TRACKER System.

In addition, only connect the 35360A TRACKER Display to an

AC power outlet employing a third wire safety ground.

Always turn the 35360A TRACKER Display's power off

prior to connecting or disconnecting the interface cable.

Failure to do so may result in an electrical shock.

The Model 35360A TRACKER Display and 35300A

Detector contain no internal operator serviceable parts.

Disassembly of either unit could result in electrical shock.

WARNING

for bias error indication.

Display only)

error indication.

Display will require repair or replacement

and interface cable)

interface cable)