LaserSpeed 8000-3 Instruction Handbook

LaserTrak Software

Part No. 93463 / Drawing No. 0921-01561

Page 111 of 221

Revision A (Sep 2007)

Low Speed Pulse Config The low speed pulse output defaults to outputting pulse according to the Low

Speed Pulse Rate set in the standard Configuration Tab. The pulse output can
be configured to output gauge status if the low speed pulse output is not
needed. The choices are shown in the table.

Length Threshold A

Indicates when the current length has exceeded the Length Threshold
A set point. This output is reset when the Index Pulse resets. See the

Index Pulse/Length Thresholds

section for details about length

thresholds.

Length Threshold B

Indicates when the current length has exceeded the Length Threshold
B set point. This output is reset when the Index Pulse resets. See the

Index Pulse/Length Thresholds

section for details about length

thresholds.

Advanced Operating Parameters

Direction Inversion
Switch

Changes the function of the Direction Digital input.

When set to Normal, a low voltage on the Direction Input (or leaving it
disconnected) will cause the gauge to measure a positive velocity and the
length to count up, and a high (5-24V) voltage on the Direction Input will cause
the gauge to measure a negative velocity and the length to count down.

When set to Reverse, the polarity of the digital input is reversed, so a low will
cause the gauge to measure a negative velocity, and a high (5-24V) will cause
the gauge to measure a positive velocity.

Changing the direction will also change the phase of the pulse outputs,
causing a connected quadrature counter to count in the opposite direction.

Advertising