Nanalog zero scale velocity, Efilter range enables, Smaterial present input mode – Laser beta lasermike LS8000-3 User Manual

Page 132

Advertising
background image

LaserSpeed 8000-3 Instruction Handbook

RS-232/RS-422 Communication

Part No. 93463 / Drawing No. 0921-01561

Page 132 of 221

Revision A (Sep 2007)

N

Analog Zero Scale Velocity

N<CR>

Read

Response

Format:

<N>

N<N><CR>

Set

<N>

is the velocity value * 1000 corresponding to the zero scale output of the 0-2V analog

output. The Analog Output is described in detail in the

Analog Output Voltage

section.

Note: Changing this setting will cause a reset, will reset the indicated length, and will cause
the pulse outputs to be momentarily interrupted. Do not change this setting during normal
operation.

E

Filter Range Enables

E<CR>

Read

Response

Format:

<N>

E<N><CR>

Set

<N>

is an integer that should normally be set to 31.

This setting should normally be left at 31 and should not be modified. The LS8000 will not
work properly if this setting is set incorrectly. LaserTrak uses this value to configure the
minimum and maximum measurable speeds. Only change this setting if specifically
instructed to do so by a Beta LaserMike Service Engineer.

S

Material Present Input Mode

S<CR>

Read

Response

Format:

<N>

S<N><CR>

Set

<N>

is an integer from 0 to 2 indicating the Material Present Input Mode.

Value Description

0

Use Internal Material Present

1

External Material Present (Active Low)
MP High Low transition starts a measurement
MP Low High transition ends measurement

2

External Material Present (Active High)
MP Low High transition starts a measurement
MP High Low transition ends measurement

Note: To disable the Material Present feature and have Material Present on all the time, set
the Material Present Input Mode to External Active Low and leave the Material Present input
disconnected.

Advertising