LS8000-3E/LS8000-3X Breakaway Cable Pinout

LaserSpeed breakaway cables have 39-pin military connectors on both ends
(male on one end and female on the other) and have pinouts identical to the
standard gauge cable.

39-Pin Male

39-Pin

Female

Description

1

1

RS232 Transmit (from LS8000-3 to host)

2

2

RS232 Receive (from host to LS8000-3)

3

3

Phase A True – High Speed Output (RS-422 Drivers)

4

4

Phase A False – User Scaleable (5-24V Output)

5

5

Phase A False – High Speed Output (RS-422 Drivers)

6

6

Phase B True – User Scaleable (5-24V Output)

7

7

Phase B True – High Speed Output (RS-422 Drivers)

8

8

Phase B False – User Scaleable (5-24V Output)

9

9

Phase B False – High Speed Output (RS-422 Drivers)

10

10

Material Present Input (5-24V Input)

11

11

Signal Ground for Inputs/Outputs/Serial

12

12

Power Ground for 24V Input

13

13

Power Ground for 24V Input

14

14

Measurement Direction Input (5-24V Input)

15

15

Phase A True – User Scaleable (5-24V Output)

16

16

Laser Interlock (Connect to Signal Ground to Turn On Laser)

17

17

Shutter Control (Connect to Signal Ground to Open Shutter)

18

18

Length Reset Input (5-24V Input)

19

19

Signal Ground for Inputs/Outputs/Serial

20 20

User V

IN

– Voltage input for Isolated Pulse Outputs (5 to 28V DC). The

voltage supplied will be the voltage level of the pulse outputs supplied by the
LS8000-3. If a Voltage is not supplied, the pulse outputs will be
approximately 4.5V.

22

22

Signal Ground for Inputs/Outputs/Serial

24

24

Index Pulse True - User Scaleable (5-24V Output)

25

25

Index Pulse False - User Scaleable (5-24V Output)

26 26

24V

Fused

Input

27 27

24V

Fused

Input

28

28

RS422 Transmit + (from LS8000-3 to host)

29

29

RS422 Transmit – (from LS8000-3 to host)

30

30

RS422 Receive + (from host to LS8000-3)

31

31

RS422 Receive – (from host to LS8000-3)

32

32

Analog Output Voltage

33

33

Analog Output Ground

34

34

Measurement Synchronization Input +

35

35

Measurement Synchronization Input -

36 36

Ethernet

TX+

37 37

Ethernet

RX+

38 38

Ethernet

TX-

39 39

Ethernet

RX-