Appendix B: Examples

B–2

Option 01 VXI Interface Module User Manual

This response indicates that the Power On bit of the Standard Event Status
register is set. This indicates that this is the first time an *ESR? query has
been sent to the Option 01 since power was applied to the module.

H Check the contents of the Standard Event Status register again.

WRITE

*ESR?

READ

000

The content of the Standard Event Status register is equal to zero because
this register is read destructively. That is, it is set to zero after its contents are
reported following an *ESR? query.

H Enable VXI TTL trigger 1 to be pulsed low for 3

µ

sec after a channel is

closed on any of the relay modules.

WRITE

output:ttltrg1:state on

H Check to see if VXI TTL trigger 1 is enabled.

WRITE

output:ttltrg1:State?

READ

1

A response of 1 indicates that VXI TTL trigger 1 is enabled. A response of 0
indicates that this trigger is disabled.

H Close channel 3 in section 5 of the VX4320 Module.

WRITE

close (@m1(3!5))

H Configure the VX4330 as a 240-to-1 1-wire scanner.

WRITE

route:conf:join m2,(1:6)

WRITE

route:conf owire,m2,(1:6)

WRITE

route:close:mode scan,m2,(1:6)

H Close channel 1 of section 1 of the VX4330.

WRITE

route:close (@m2(1!1))

H Close channel 40 of section 2 of the VX4330.

WRITE

route:close (@m2(40!2))

Note that channel 1 of section 1 will be opened before channel 40 of section
2 is closed because sections 1 and 2 have been joined and both sections have
been set to a close mode of “scan”.

H Close channels 1 through 10 on the VX4350 Module.

WRITE

close (@m3(1:10))