Command Syntax

Option 01 VXI Interface Module User Manual

3–9

For commands directed to a VX4372 or VX4374 which has two scanner
sections:

<section_list>

Sections Specified

(1)

Section 1

(1,2)

Sections 1 and 2

(1:2)

Sections 1 and 2

error/event queue When the command parser detects a syntax error or data range
error, it places an error message describing the error in the Error/Event queue.
Bit 2 of the Status Byte Register is set to indicate that this queue is not empty.
Bit 5 of the Standard Event Status Register (the Command Error bit) is set if the
parser detects a syntax error. Bit 4 of the Standard Event Status register (the
Execution Error bit) is set if the parser detects a numeric argument that is out of
range. When a SYSTem:ERRor? query is received, an error message is dequeued
from the Error/Event queue and placed in the Output queue.