Appendix B: Examples

Option 01 VXI Interface Module User Manual

B–3

H Open channels 11 through 20 on the VX4350 Module.

WRITE

open (@m3(11:20))

H Query the close state of channels 1 through 20 on the VX4350 Module.

WRITE

close? (@m3(1:20))

READ

1 1 1 1 1 1 1 1 1 1 0 0 0 0 0 0 0 0 0 0

A query response of 1 indicates that the corresponding channel is closed, and
a response of 0 indicates that the channel is open.

H Query the open state of channels 1 through 20 on the VX4350 Module.

WRITE

open? (@m3(1:20))

READ

0 0 0 0 0 0 0 0 0 0 1 1 1 1 1 1 1 1 1 1

A query response of 1 indicates that the corresponding channel is open, and a
response of 0 indicates that the channel is closed.

H Join the two sections of the VX4372 Module.

WRITE

route:conf:join m4,(1,2)

H Close channel 10 in section 1 of the VX4372 Module.

WRITE

close (@m4(10!1))

H Disjoin the two sections of the VX4374 Module.

WRITE

route:conf:disjoin m5

H Close channel 12 of section 2 of the VX4374.

WRITE

route:close (@m5(12!2))

H Query the close state of the 12 channels in section 2 of the VX4374.

WRITE

route:close? (@m5(1!2:12!2))

READ

0 0 0 0 0 0 0 0 0 0 0 1

Note that this query can also be specified as

route:close? (@m5(13:24))

H Open all relays in section 4 of the VX4380 Module.

WRITE

route:open (@M6(1!1!4:4!16!4))

H Close the relay that joins row 3 to column 12 in section 4 of the VX4380

Module.

WRITE

route:close (@m6(3!12!4))