Trigger – Tektronix 070-9180-01 User Manual

Page 79

Advertising
background image

Command Descriptions

Option 01 VXI Interface Module User Manual

3–55

TRIGger

TRIGger

[:SEQuence]

:COUNt

N/A

1

The count specified in this command must be between 1 and 65535.

ABORt, ROUTe:SCAN
TRIGger[:SEQuence]:SOURce
TRIGger[:SEQuence][:IMMediate]
TRIGger[:SEQuence]:DELay

This command specifies the number of times to sequence through an entire scan
sequence.

Command

Response

route:scan

(@m1(1,2),m2(1:64))

Define a scan sequence consisting of relays 1 and 2 on relay
module m1 and relays 1 through 64 on module m2.

trigger:sequence:count

2

Define the number of times to sequence through the scan list
before setting the state of the trigger subsystem back to the
idle state.

trigger:sequence:

source:immediate

Sequence through the scan list without waiting for a trigger
event after an INITiate[:Immediate] command is received.

initiate:immediate

Initiate the scan sequence. Sequence through the entire scan
list two times.

TRIGger

[:SEQuence]

:DELay

N/A

Command Syntax

Query Syntax

*RST Value

Limits

Related Commands

Description

Examples

Command Syntax

Query Syntax

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com