Tektronix 070-9180-01 User Manual

Page 81

Advertising
background image

Command Descriptions

Option 01 VXI Interface Module User Manual

3–57

ABORt, ROUTe:SCAN
TRIGger[:SEQuence]:SOURce
TRIGger[:SEQuence]:COUNt
TRIGger[:SEQuence]:DELay

Sequence to the next relay in an initiated scan list without waiting for a trigger
event and without waiting the time specified in a TRIGger[:SEQuence]:DELay
command.

Command

Response

route:scan

(@m1(1,2),m2(1:64))

Define a scan sequence consisting of relays 1 and 2 on relay
module “m1” and relays 1 through 64 on module m2.

trigger:sequence count

2

Define the number of times to sequence through the scan list
before setting the state of the trigger subsystem back to the
idle state.

trigger:sequence:delay

0.5

Wait 0.5 seconds after a trigger event is detected before
opening the current relay in a scan list and closing the next
relay in the list.

trigger:sequence:

source:ttltrg2

Sequence to the next relay in a scan list when VXI TTL trigger
2 is asserted.

initiate:immediate

Initiate the scan sequence. Sequence through the entire scan
list 2 times. Wait 0.5 seconds after VXI TTL trigger 2 is
asserted before sequencing to the next relay in the list.

trigger:sequence:

immediate

Sequence to the next relay in the scan list without waiting for
VXI TTL trigger 2 to be asserted. Do not wait 0.5 seconds to
sequence to the next relay.

TRIGger

[:SEQuence]

:SOURce BUS | HOLD | IMMediate | TTLTrg<n>

N/A

IMMediate

N/A

ABORt, ROUTe:SCAN
TRIGger[:SEQuence][:SOURce]
TRIGger[:SEQuence][:IMMediate]

Related Commands

Description

Examples

Command Syntax

Query Syntax

*RST Value

Limits

Related Commands