Getting Started

Option 01 VXI Interface Module User Manual

1–3

VXI interface &
slave controller

daughter board

VXI backplane
local bus

Backplane

connectors

Daughter board

connectors

Front panel

connectors

Figure 1–2: Typical SurePath Configuration

Each relay module that the Option 01 Module controls has a fuse that limits the
amount of current that the module can draw from the VXI backplane +5 V power
pins. This fuse protects the module in case of an accidental shorting of the power
bus or any other situation where excessive current might be drawn.

The Option 01 Module monitors the state of the fuses of each of the relay
modules that it controls. If a fuse opens on one of these modules, the Option 01
Module will assert SYSFAIL* on the VXIbus.

If the +5 V fuse on the module on which the Option 01 Module is installed
opens, the VXIbus Resource Manager will be unable to assert SYSFAIL
INHIBIT on the Option 01 to disable SYSFAIL*.

If a +5 V fuse opens, remove the fault before replacing the fuse. Replacement
fuse information is given in the Specifications section of the user manual for the
appropriate relay module.

Built-in Test Equipment (BITE) is provided by extensive self tests that are
automatically invoked on power-on and can also be invoked on command.
Circuitry tested includes the CPU and all memory

,

and the relay control circuitry

on each relay module controlled by the Option 01.

Fuses

BITE (Built-in Test

Equipment)