Ex ia-version, Electrical connection, 46 vegapuls 56v – VEGA VEGAPULS 56V User Manual

Page 46: Exd-connection housing, Cover lockings

Advertising
background image

46

VEGAPULS 56V

- +

2 1

Shield

Supply: 20 … 36 V DC, VBUS

1

2

- +

ESC

+

-

OK

1

2

C

5

6

7

8

4

3

VBUS

Commu-
nication

+ -
4...20mA

Display

1

2

C

5

6

7

8

4

3

EEx d-connection housing

(opening in Ex-area not permitted)

Adjustment module and indicating
terminal box

(opening in Ex-area permitted)

Exd-connection housing

1

/

2

“␣ NPT EEx␣ d

diameter of the
connection cable
to the Exd-
connection
housing
3,1…8,7␣ mm
(0,12…0,34␣ inch)

Exd-proof seal to the Exd-
connection housing

1

/

2

“␣ NPT EEx␣ d

diameter of the
connection cable
3,1…8,7␣ mm
(0,12…0,34␣ inch)

Exd-version (loop-powered with pressure-tight encapsulated terminal box)

Cover lockings

ESC

+

-

OK

-

+

1

2

C

5

6

7

8

4

3

VBUS

Commu-
nication

+ -
4...20mA

Display

1

2

C

5

6

7

8

4

3

Voltage supply and
digital measuring signal

To the indicating instrument in the
sensor cover or to the external indicat-
ing instrument VEGADIS 50

M20 x 1,5
(diameter of the
connection cable
5…9␣ mm)

Ex ia-version

Electrical connection

Voltage supply and
digital measuring signal

Spring terminals (max.
2,5␣ mm

2

cross-section area

of conductor)

Spring terminals (max.
2,5␣ mm

2

cross-section area

of conductor)

Sockets for connection of
VEGACONNECT 2 (communi-
cation sockets)

Advertising