VEGAPULS 56V

49

When the signal conditioning instruments are
exceptionally not pre-configured, start now
with the following section "

Configuration“ and

continue then with the adjustments in section
"Parameter adjustment“.

Configuration

Create new measurement loop

• Choose the menu "Configuration/Measure-

ment loop/New

“ and you are in the menu

window "Create new measurement loop -
Application“.

• Choose the parameter ("Level measure-

ment“, "gauge“ or "distance“) and the sen-
sor type ("Pulse-Radar“).

• Click to "Continue“.

Set-up

6.2 Adjustment with the PC on

VEGAMET

For connection of the PC to the signal condi-
tioning instrument the interface converter
VEGACONNECT 2 is required. The PC com-
municates via the interface converter with the
signal conditioning instrument and with the
connected sensor (s).

The individual adjustment steps are marked
in the following with a dot:
• Choose …
• Start …
• Click to …

Now start:
• Connect the standard plug of VEGACON-

NECT 2 (9-pole) with the interface COM 1
or COM 2 of your PC.

• Insert the two small pin plugs of VEGA-

CONNECT 2 into the CONNECT-sockets
on the front side of the signal conditioning
instrument.

• Now switch on the power supply of the

signal conditioning instrument.

After approx. 1␣ …␣ 2 minutes (self-check) the
measuring system is generally in operating
condition and displays measured values.

• Start the adjustment software VVO on your

PC.

Instruction before starting the
configuration:

The signal conditioning instruments are al-
ready configured according to the sensor
type ordered with the signal conditioning
instrument.

Generally you will therefore use a pre-config-
ured signal conditioning instrument. In the
following menu "

Configuration“, normally no

adjustments are necessary and you can
directly choose the menu "Parameter adjust-
ment“ on page 51.